SRT 100 Product Specifications

The SRT 100 has been developed to provide the following unique operational advantages:

TOPEX technology is based on several needs requirements. The most prevalent is the need for a current dedicated skin cancer treatment device. There have been little to no new developments in superficial radiation therapy systems in recent years. The earlier devices were developed to provide therapeutic services for treating a number of cancer conditions and have been replaced by other modalities and/or specialized procedures. However, the primary use for these older systems today is skin cancer treatment. Consequently, many systems still in operation are antiquated by today's standards in terms of system operation, long set-up procedures, large space requirements, parts availability and ongoing costly maintenance to insure continued operation and that product safety requirements are met for both the patient and clinician.



Product Specifications

Generator

  • Type - Constant Potential HV
  • Input Line - 120 - 230 VAC

X-Ray Tube

  • Metal Ceramic
  • Water -cooled
  • Tungsten Target
  • End grounded
  • Rating - 120kV/10mA, 40kV/30mA,
    1200 watts continuous dissipation
  • Focal Spot Size - 1mm x 1mm

Base Unit Assembly

  • Base Space Requirements - 29" by 30"
  • Vertical Travel - up to 68"
  • Horizontal Arm Travel - 17" to 61" from center
  • X-Ray Tube Movement - V&H 180 degrees
  • Integrated Modular Components - Input power, HV Generator, Heat Exchanger

Operator Control Console

  • Can be located up to 100' (30meters) from base unit
  • Service mode for system set-up and calibration - key entry

Two Treatment Techniques

  • 100kV @ 8mA, 2.1 mm Al. HVL
  • 70kV @ 10mA, 1.3mm Al. HVL
  • X-Ray output is 460 cGy @ 15 cm SSD

Automatic Filter Change (Patent pending)

  • 2.1 mm Al. HVL
  • 1.3 mm Al. HVL
  • Pb X-ray Block

Automatic Warm Up procedures

  • Automatically activated from time of last exposure
  • Four pre-programmed sequences
  • Pb lead blocker automatically placed over x-ray tube port

RAD Check

  • Direct radiation measurement of output
  • Pre-treatment verification

System Weight

  • 350 lbs. (160 kg)

Accessories

Standard Size Treatment Applicators

  • 1.5cm, 3cm and 5cm Diameter @ 15cm SSD

Optional Size Treatment Applicators

  • 1.0 cm, 2 cm, 2.5 cm, 4.0 cm Diameter @ 15cm SSD
  • 10 cm Diameter @ 25 SSD

Replaceable Safety Contact Shields

  • Applicator size specific
  • Visibility of treated area

Base Unit Dimentions
Base Unit Dimentions